| Model Number |
5820012040 |
| Medical Device Problem Codes |
Flaked (1246); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a lumber surgical procedure, the bur broke, the procedure was completed successfully without a delay; no adverse consequences or medical intervention were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A follow up report will be filed once the quality investigation is complete.Awaiting confirmation if device is available for return.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: it was initially reported, that the bur broke.Upon evaluation, the diamond grit fragments fell off the head of the bur.H6: the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported, that during, a lumber surgical procedure.The bur broke.The procedure was completed successfully without a delay.No adverse consequences or medical intervention were reported.It was further reported, that following evaluation of the returned bur, it not break.The diamond grit fragments fell off the head of the bur.
|
| |
|