| Catalog Number |
46535BER |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow-up report will be submitted once the evaluation has been complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during an embolization procedure, the catheter detached within the patient.The physician had acquired retrograde femoral artery access and had negotiated the patient's iliac bifurcation with a guidewire and the 4f catheter.During over-the-wire catheter exchange attempt, within the patient's tortuous and calcified left internal iliac artery the catheter tip detached.The tip remained on the guidewire and was successfully removed along with the catheter.Additional medical intervention was not necessary.The primary procedure site was abandoned due to sheath vessel tortuosity and calcification.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect medical device was returned for evaluation.The device was examined visually.The complaint is confirmed.The root cause could not be determined however, it is likely that significant force was applied to the device during clinical use.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.
|
| |
|