• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAYENNE MEDICAL QUATTRO LINK ANCHOR 4.5MM; FASTENER, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAYENNE MEDICAL QUATTRO LINK ANCHOR 4.5MM; FASTENER, FIXATION Back to Search Results
Model Number N/A
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during surgery that two anchors were bent and not disengaging properly.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 3006108336 - 2022 - 00036.Concomitant medical products: part # cm-9324st 2.9mm gl hard bone drill lot # 65415347 qty 1.Part # cm-9145 quattro link anchor 4.5mm lot # 81553-2 qty 1.Part # cm-9155 quattro link anchor 5.5mm lot # 67346-3 qty 2.Part # cm-9145 quattro link anchor 4.5mm lot # 65415424 qty 2.Part # cm-9155 quattro link anchor 5.5mm lot # 67346-3 qty 1.Products have been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Products have been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified the bent ¿cleats¿.Both devices inserter prongs are bent at roughly 45 degrees, and one has the prongs pushed together.The inserters¿ adjustment knobs and trigger mechanisms function smoothly and extend and retract the inserter shaft.It is noted that the anchor was not returned for evaluation.Medical records were not provided.Device history records were reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.Corrected, event is no longer reportable; no serious injury or harm to the patient reported for this event or prior events with same or similar products.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUATTRO LINK ANCHOR 4.5MM
Common Device Name
FASTENER, FIXATION
Manufacturer (Section D)
CAYENNE MEDICAL
16597 n 92nd street
scottsdale AZ 85260
Manufacturer (Section G)
CAYENNE MEDICAL
16597 n 92nd street
scottsdale AZ 85260
Manufacturer Contact
jennifer rapsavage
56 e. bell dr
warsaw, IN 46582
5745260384
MDR Report Key16018740
Report Number3006108336-2022-00041
Device Sequence Number6894308
Product Code MBI
UDI-Device Identifier00887868270192
UDI-Public(01)00887868270192(17)260806(10)81553-2
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K122314
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/06/2026
Device Model NumberN/A
Device Catalogue NumberCM-9145
Device Lot Number81553-2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2022
Supplement Date Received by Manufacturer01/09/2023
03/31/2023
Initial Report FDA Received Date12/20/2022
Supplement Report FDA Received Date02/03/2023
04/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Patient SexFemale
-
-