This report has been identified as b.Braun internal report number (b)(4).Received two pictures from the customer of the complained introcan safety 3 pur 18g 1.3x32mm-sa.As the affected sample was not provided and the pictures received were poor quality, proper visual assessment on the pictures was unable to be done.Unable to review the device history record for batch number 21l06g892 as the batch number provided does not exist in sap mya, further evaluation is not possible and the exact cause of the failure could not be determined.The in-line test equipment is subject of a frequent calibration and a regular verification related to its proper function.Herewith potential malfunctions of the systems would be detected in-time and would be mitigated.Beside the automated 100% in-line test equipment, independent in-process quality controls and final controls on a random sample basis will be conducted by different teams on a regular basis within the production process.Herewith a systematic product defect would be detected.Tear off capillary defect most likely not appear to be attributed by manufacturing process as the defect is able to be detected by the in-line vision system.Damages induced after assembly process is not possible since the catheter had been protected with the protective cap.Cause : defect due to wrong handling (referring to application error or off-label use) justification: not confirmed.Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
|