B. BRAUN MELSUNGEN AG INTROCAN® SAFETY; I.V. SAFETY CATHETER Back to Search Results
Model Number 4251601-020
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 10/16/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report has been identified as b.Braun internal report number (b)(4).Received two pictures from the customer of the complained introcan safety 3 pur 18g 1.3x32mm-sa.As the affected sample was not provided and the pictures received were poor quality, proper visual assessment on the pictures was unable to be done.Unable to review the device history record for batch number 21l06g892 as the batch number provided does not exist in sap mya, further evaluation is not possible and the exact cause of the failure could not be determined.The in-line test equipment is subject of a frequent calibration and a regular verification related to its proper function.Herewith potential malfunctions of the systems would be detected in-time and would be mitigated.Beside the automated 100% in-line test equipment, independent in-process quality controls and final controls on a random sample basis will be conducted by different teams on a regular basis within the production process.Herewith a systematic product defect would be detected.Tear off capillary defect most likely not appear to be attributed by manufacturing process as the defect is able to be detected by the in-line vision system.Damages induced after assembly process is not possible since the catheter had been protected with the protective cap.Cause : defect due to wrong handling (referring to application error or off-label use) justification: not confirmed.Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
 
Event or Problem Description
As reported by the user facility (translation of user facility information by bbm sales organization in colombia: "broken device." "patient who consulted a health institution for abdominal pain on (b)(6) 2022, who was given an intravenous catheter for medication.The catheter is removed from the patient for discharge.Patient re-consults for pain at the catheter insertion site and poor mobility.The patient reported "i feel like i'm squeezing".With ultrasound the attending physician identifies intramuscular foreign body with thrombosis and requires referral to a vascular physician for foreign body removal.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN® SAFETY
Common Device Name
I.V. SAFETY CATHETER
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen, hessen 34212
GM  34212
Manufacturer (Section G)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen 34212
Manufacturer Contact
jonathan severino
861 marcon blvd.
allentown 18109
MDR Report Key16018952
Report Number9610825-2020-00541
Device Sequence Number5896596
Product Code DQR
UDI-Device Identifier04046964478673
UDI-Public(01)04046964478673(17)250501(10)20E06G8911
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K982805
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4251601-020
Device Catalogue Number4251131-04
Device Lot Number21L06G892
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2022
Initial Report FDA Received Date12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-