• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. SILK SUTURE 30"(75CM) 4-0 BLK; SUTURE, NONABSORBABLE, SILK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. SILK SUTURE 30"(75CM) 4-0 BLK; SUTURE, NONABSORBABLE, SILK Back to Search Results
Model Number K952H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Events reported via: 2210968-2022-10494.
 
Event or Problem Description
It was reported that a patient underwent a dental procedure on (b)(6) 2022 and suture was used.During the procedure, the needle was detached from the suture during suturing.This happened to two products in a row.The products were not control release needles and used on oral.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 1/18/2023.H6 component code: g07002 no device problem found.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection evaluation was conducted on the returned device.Visual analysis of the returned sample determined that three paper lids and one detached needle that pertains to product code k952h were returned to ethicon for analysis.As per the visual inspection of the needle, the swage and attachment area was noted to be as expected.Marks on the body needle that appears to be by a surgical instrument could be observed.No suture remnant could be observed into the barrel hole.The suture was not received for evaluation to determine the assignable cause.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what caused the reported complaint.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: visual analysis of the returned sample determined that it was received one open box with eighteen unopened samples that pertain to the product code k952h.As per the sampling plan, a visual inspection was performed on thirteen samples, the packets were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand no anomalies were observed during the evaluation.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 2/10/2023.H3 investigational summary: the product received for analysis was identified as product code k952h.Visual inspection and metallurgical analysis were performed on the returned product.A fracture was observed at the suture attachment of the needle.One side of the needle was received, the mating fracture surface was not provided for this evaluation.The needle was noted with marks that appears to be by surgical instrument.A scanning electron microscope was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation of the sample revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.This was a ductile fracture.The evidence of this examination indicates that the breakage occurred at the attachment area of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILK SUTURE 30"(75CM) 4-0 BLK
Common Device Name
SUTURE, NONABSORBABLE, SILK
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km. 8.3
san lorenzo PR
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key16019288
Report Number2210968-2022-10493
Device Sequence Number12486327
Product Code GAP
UDI-Device Identifier10705031044760
UDI-Public10705031044760
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberK952H
Device Catalogue NumberK952H
Device Lot NumberSDBEAH
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/24/2022
Supplement Date Received by Manufacturer01/03/2023
02/03/2023
Initial Report FDA Received Date12/20/2022
Supplement Report FDA Received Date01/18/2023
02/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-