Model Number MAJ-2138
Medical Device Problem Codes Break (1069); Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device manufacturer date: the manufacture date cannot be identified at this time since the lot was not provided.If the lot becomes available, the manufacturer date will be updated part of the device was returned to olympus, the customer only returned the connecting tube lot number unknown for evaluation.A visual inspection on the as is received condition of the device; noted that the device was not returned in its original packaging.It was confirmed that one of the gray clips on the connector was missing and not returned for evaluation.The tubing appears to have no signs of damage as there are no punctures or kinks identified.Several scratches were observed on the metal connector.Unable to perform any functional tests as one of the clips is missing.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event or Problem Description
The customer reported to olympus that during initial use the connecting tube was broken and cannot be connected to the channel connector in the reprocessing basin.There were no reports of patient harm associated with this event.
 
Additional Manufacturer Narrative
This report is being submitted to provide additional information based on the legal manufacturer¿s final investigation.Since the manufacturing date of the device is unknown, the device history record was unable to be reviewed, however, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the device investigation, the cause of the reported ¿lock lever of the cleaning tube connector came off¿ issue was confirmed.The root cause of the damaged lock levers could not be determined, however, this issue was found to likely be the result of an external load applied to the cleaning tube, making it impossible to connect.An external load on the irrigation tube may be caused by the user applying force in the wrong direction.The event can be prevented by following the instructions for use which state: ¿abnormalities can be detected by conducting the following inspections: 9 preparation and inspection: 2 move the lock levers of the reprocessor side connector to make sure that they function properly and are not broken¿.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECTING TUBE
Common Device Name
ENDOSCOPE REPROCESSOR ACCESSORIES
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581-4003
5082077661
MDR Report Key16022002
Report Number9610595-2022-05855
Device Sequence Number6922494
Product Code FEB
UDI-Device Identifier04953170404184
UDI-Public04953170404184
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K190969
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMAJ-2138
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2022
Supplement Date Received by Manufacturer01/05/2023
Initial Report FDA Received Date12/21/2022
Supplement Report FDA Received Date01/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-