CARDIAC SURGERY MFG KERKRADE CUSTOM TUBING PACK; TUBING, PUMP, CARDIOPULMONARY BYPASS Back to Search Results
Model Number M331107A
Device Problems Material Separation (1562); Material Deformation (2976); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2022
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Medtronic received information that during use of a custom tubing pack it was reported that when the perfusionist pulled on the arterial line, the connector assembly came off.The custom tubing pack was replaced with a custom tubing pack from the same lot to complete the procedure.Medtronic received additional information that this custom tubing pack was not used as the issue was detected before patient use, after priming.There was no patient impact associated with this event.Medtronic received additional information that it was not the oxygenator arterial line connector that got disconnected.It was the oxygenator temperature probe that came loose.
 
Manufacturer Narrative
Correction b5: medtronic received additional information that there was no assembly problem with the custom tubing pack prior to the procedure, this was reported in error.There was only the reported issue with the temperature probe of the oxygenator at the end of the procedure.It was reported that the perfusionist forced the temperature probe and saw the glue crack.Correction h6: imf f27 added.Additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event did not and does not meet the reporting requirements stipulated in 21 cfr 803.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM TUBING PACK
Type of Device
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
CARDIAC SURGERY MFG KERKRADE
valkenhuizerlaan 16a
kerkrade 6466 ND
NL  6466 ND
Manufacturer (Section G)
CARDIAC SURGERY MFG KERKRADE
valkenhuizerlaan 16a
kerkrade 6466 ND
NL   6466 ND
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16022057
MDR Text Key308381120
Report Number6000033-2022-00019
Device Sequence Number1
Product Code DWE
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K171979
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/15/2024
Device Model NumberM331107A
Device Catalogue NumberM331107A
Device Lot Number225235965
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/28/2022
Date Device Manufactured10/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-