• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INSERT #5 11MM; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INSERT #5 11MM; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 5531-G-511-E
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Pain (1994); Fibrosis (3167); Swelling/ Edema (4577)
Date of Event 05/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
Patient reported having a right knee manipulation under anesthesia due to pain and swelling on (b)(6) 2022.
 
Event or Problem Description
Patient reported having a right knee manipulation under anesthesia due to pain and swelling on may 21, 2022.Update: spoke with patient who stated that he will be seeing his doctor on 14 january 2023 to determine if a revision will be done.Patient states he is "leaning towards having the revision done as he is experiencing back and bilateral hip pain".Update: date of surgery (b)(6) 2022 per operative report.
 
Additional Manufacturer Narrative
Reported event: an event regarding rom involving an unknown triathlon knee was reported.The event was confirmed via clinician review of the provided medical records.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.-clinician review: a review of the provided medical information by a clinical consultant indicated: conclusion/assessment: the patient had a poor knee range of motion postoperatively which led to mua and arthroscopic lysis of adhesions.He remains dissatisfied and is considering a revision surgery.Event confirmation: poor postoperative range of motion, mua and lysis of adhesions via an arthroscope can be confirmed.Root cause: a root cause cannot be determined with the records provided." -product history review: could not be performed as the device lot details were not provided.-complaint history review: could not be performed as the device lot details were not provided.Conclusions: it was reported that the patient underwent medical intervention to address range of motion issues.The event was confirmed through clinician review of the provided medical records.The exact cause of the event could not be determined because insufficient information was provided.Further information such as device-identifying information (catalog number and lot code), pathology reports, pre- and post-operative x-rays as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Additional Manufacturer Narrative
Reported event: an event regarding rom involving a triathlon insert was reported.The event was confirmed via clinician review of the provided medical records.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.Clinician review: a review of the provided medical information by a clinical consultant indicated: "conclusion/assessment: the patient had a poor knee range of motion postoperatively which led to mua and arthroscopic lysis of adhesions.He remains dissatisfied and is considering a revision surgery.Event confirmation: poor postoperative range of motion, mua and lysis of adhesions via an arthroscope can be confirmed.Root cause: a root cause cannot be determined with the records provided." product history review: could not be performed as the device lot details were not provided.Complaint history review: could not be performed as the device lot details were not provided.Conclusions: it was reported that the patient underwent medical intervention to address range of motion issues.The event was confirmed through clinician review of the provided medical records.The exact cause of the event could not be determined because insufficient information was provided.Further information such as device lot code, pathology reports, pre- and post-operative x-rays as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.Catalog numbers and lot codes of other devices listed in this report: 5517f502 triathlon p/a cr beaded #5r lot unknown, 5536b500 tritanium bplate triathlon s5 lot unknown, 5556-l-360 tritanium patella-symmetric lot unknown.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
 
Event or Problem Description
Patient reported having a right knee manipulation under anesthesia due to pain and swelling on (b)(6), 2022.Update: spoke with patient who stated that he will be seeing his doctor on (b)(6) 2023 to determine if a revision will be done.Patient states he is "leaning towards having the revision done as he is experiencing back and bilateral hip pain".Update: date of surgery (b)(6) 2022 per operative report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X3 TRIATHLON CS INSERT #5 11MM
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood, NJ WD6 3-SJ
UK   WD6 3SJ
2018315000
MDR Report Key16022596
Report Number0002249697-2022-01872
Device Sequence Number5884123
Product Code KRO
UDI-Device Identifier07613327336450
UDI-Public07613327336450
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K172634
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number5531-G-511-E
Device Catalogue Number5531-G-511-E
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/06/2022
Supplement Date Received by Manufacturer02/08/2023
05/05/2023
Initial Report FDA Received Date12/21/2022
Supplement Report FDA Received Date02/27/2023
05/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
Patient SexMale
Patient Weight96 KG
-
-