INTEGRA LIFESCIENCES MANSFIELD PROG VALVE CLYNDRICAL W PRECHA; CHPV Back to Search Results
Model Number 823110
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Decreased Level of Consciousness (4591)
Date of Event 11/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
2 of 2 reports (same patient, different valves).Other mfg report number: 3013886523-2022-00579.A facility reported a hakim valve was implanted on (b)(6) 2022 in a 58 year old male patient.It did not function properly and was replaced on (b)(6) 2022 with a new valve (id 823110).The second valve presented with the same failure and was explanted on (b)(6) 2022.It seems that there were some debris in the valves.Consequences for the patient: drop in consciousness, longer hospital stay.The patient status is: recovered.
 
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The hakim valve was returned for evaluation: dhr - lot 5822407, conformed to the specifications when released to stock.Failure analysis - the valve was visually inspected: no defects were noted.The valve passed the tests for programming, occlusion, leaks and pressure.No root cause could be determined as the technician could not confirm any problems with the valve at the time of investigation.The possible root cause for the issue reported by the customer, could be due to biological debris and protein build up interfering with the valve mechanism at the time of investigation no functional issues were noted with the valve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROG VALVE CLYNDRICAL W PRECHA
Common Device Name
CHPV
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key16022840
Report Number3013886523-2022-00580
Device Sequence Number17613256
Product Code JXG
UDI-Device Identifier10381780519133
UDI-Public(01)10381780519133(17)260831(10)5822407
Combination Product (Y/N)N
PMA/510(K) Number
K122118
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Model Number823110
Device Catalogue Number823110
Device Lot Number5822407
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/05/2022
Supplement Date Received by Manufacturer01/27/2023
Initial Report FDA Received Date12/21/2022
Supplement Report FDA Received Date02/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age58 YR
Patient SexMale
-
-