| Model Number |
0072360 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a presence of a silicone stopper inside the tubing of suction evacuator which was proximal to the trocar.The internal white part was perforated only on the two-third instead of whole.Therefore, it was impossible to use on craniotomy for too long as it was not perforated the full length.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported issue is inconclusive as no sample was returned for evaluation.A potential root cause for this failure could be due to "equipment not clean enough, material received already contaminated, particles in environment".It was unknown whether the device had met specifications.The product was used for treatment but it was unknown whether the product had caused the reported failure.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: a labeling review is not required because labeling could not have prevented the reported issue.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a presence of a silicone stopper inside the tubing of suction evacuator which was proximal to the trocar.The internal white part was perforated only on the two-third instead of whole.Therefore, it was impossible to use on craniotomy for too long as it was not perforated the full length.
|
| |
|