| Model Number |
4515-70-109 |
| Medical Device Problem Code |
Device Slipped (1584)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Concomitant medical devices and therapy dates, robotic assisted saw interface device, (b)(6) 2022.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
This is report 2 of 2 for the same event.It was reported that during a total knee arthroplasty procedure the robotic assisted saw interface device was very loose.It was also reported that the robotic assisted array interface device was also loose.It was reported that there were no delays in the procedure.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization reported.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The returned device was evaluated and the reported condition was not confirmed.A functional test of the clamp was performed and no issues were identified, clamp functioned as design intended.No defects were identified on the device.No defects or failures were found.The assignable root cause could not determined.
|
| |
|
Search Alerts/Recalls
|
|