• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. SET SCREW DRIVER; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. SET SCREW DRIVER; SCREWDRIVER Back to Search Results
Catalog Number 230H4001
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.Reference report 3012447612-2022-00311.
 
Event or Problem Description
It was reported that two tether set screw drivers fractured during surgery.One driver tip fractured off completely in the set screw and remained implanted in the patient, the other fractured incompletely and was retrieved.This is report two of two for this event.
 
Event or Problem Description
It was reported that two tether set screw drivers fractured during surgery.One driver tip fractured off completely in the set screw and remained implanted in the patient, the other fractured incompletely and was retrieved.This is report two of two for this event.
 
Additional Manufacturer Narrative
Additional information in h4 and h6: component, investigation type, findings, and conclusions.Device evaluation: the returned devices match the information in the complaint file and were examined.Visual inspection revealed that both devices were fractured at the tip.Lot 311982 was not fully fractured off, while lot 313772 was.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.This event could also be attributed to off-axis forces applied during use.Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SET SCREW DRIVER
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key16023546
Report Number3012447612-2022-00312
Device Sequence Number20582517
Product Code HXX
UDI-Device Identifier00887868227431
UDI-Public(01)00887868227431(10)313772
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number230H4001
Device Lot Number313772
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2022
Supplement Date Received by Manufacturer05/30/2023
Initial Report FDA Received Date12/21/2022
Supplement Report FDA Received Date05/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/28/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-