• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEGADYNE MEDICAL PRODUCTS, INC. 10FT SMOKE PENCIL UCONN; RETURN ELECTRODE CABLE, DUAL PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEGADYNE MEDICAL PRODUCTS, INC. 10FT SMOKE PENCIL UCONN; RETURN ELECTRODE CABLE, DUAL PLATE Back to Search Results
Model Number 251010J
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Complete Loss of Power (4015)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 11/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the 251010j device was returned with the smoke tube broken.In addition, an electrode tip was returned attached to the device.The device was connected to the generator and the device activates when the buttons were depressed.There were no issues on the device related to the event reported.It is possible that the damage to the tube lines could be due to an excessive pull and tubing stretched.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No lot or batch were provided, bhr could not be performed.Additional information was requested, and the following was obtained: what is the severity of the burn? (please see degrees of burns below and choose one) first degree burns are minor burns on the first layer of skin.The skin looks dry, redness, and may be swelling; no penetration or blisters second degree burn looks wet or moist.The burn site appears red, blistered, and may be swollen and painful third degree burn the burn site looks deep, whitening or blackened and charred what medical intervention was used to treat the burn? (such as salve or stitches) besides the burn, did the patient experience any adverse consequence due to the issue? are there any anticipated long-term effects from the burn or injury? answer = the burn was a first degree.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during a low anterior resection procedure that while the device was laid down and not in use it self-activated burning through the surgeons gown and burning his arm.Unknown as to what degree of burn to the surgeons arm.Another like device was used to complete the procedure.There were no adverse consequences for the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10FT SMOKE PENCIL UCONN
Common Device Name
RETURN ELECTRODE CABLE, DUAL PLATE
Manufacturer (Section D)
MEGADYNE MEDICAL PRODUCTS, INC.
4545 creek road
cincinnati OH 45242
Manufacturer (Section G)
MEGADYNE MEDICAL PRODUCTS, INC.
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key16023848
Report Number1721194-2022-00118
Device Sequence Number6923074
Product Code GEI
UDI-Device Identifier10614559105573
UDI-Public10614559105573
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K982130
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number251010J
Device Catalogue Number251010J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/21/2022
Initial Report FDA Received Date12/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-