| Catalog Number |
MB3355 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Codes |
Dysphagia/ Odynophagia (1815); Joint Laxity (4526); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a revision surgery was performed to remove two mobi-c devices because they were locked in a hyperlordotic position in flexion and extension.This is report one of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3004788213-2022-00133.
|
| |
|
Event or Problem Description
|
|
It was reported that a revision surgery was performed to remove two mobi-c devices because they were locked in a hyperlordotic position in flexion and extension at c3/4 and c4/5.The patient also had dysphagia.The poly core was protruding anteriorly and subsided at c4/5.A corpectomy with fusion was performed at c4 during the revision.This is report one of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported a revision surgery was performed to remove two mobi-c devices because they were locked in a hyperlordotic position in flexion and extension at c3/4 and c4/5.The patient also had kyphosis that led to dysphagia, but no pain.The poly core was protruding anteriorly and subsided at c4/5.A corpectomy with fusion was performed at c4 during the revision.This is report one of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information in h6: component, investigation type, findings, and conclusions.Inspection: x-rays were provided which show that both implants are in a hyperlordotic or "fish mouth" position.Migration of the poly core could not be confirmed from the provided x-rays.The devices were sent to a lab for review.The lab provided photos of the devices which show that one of the poly cores is fractured.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the devices were likely conforming when they left zimvie¿s control.Potential root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to unknown patient or surgical factors.Device usage: this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
|
| |
|
Event or Problem Description
|
|
It was reported a revision surgery was performed to remove two mobi-c devices because they were locked in a hyperlordotic position in flexion and extension at c3/4 and c4/5.The patient also had kyphosis that led to dysphagia, but no pain.The poly core was protruding anteriorly and subsided at c4/5.A corpectomy with fusion was performed at c4 during the revision.This is report one of two for this event.
|
| |
|