MAUDE Adverse Event Report: LDR MÉDICAL MOBI-C IMPLANT "M" STANDARD 13X15 H5 US; MOBI-C CERVICAL DISC PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LDR MÉDICAL MOBI-C IMPLANT "M" STANDARD 13X15 H5 US; MOBI-C CERVICAL DISC PROSTHESIS Back to Search Results
Catalog Number MB3355
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Codes Dysphagia/ Odynophagia (1815); Joint Laxity (4526); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 12/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed to remove two mobi-c devices because they were locked in a hyperlordotic position in flexion and extension.This is report one of two for this event.
 
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3004788213-2022-00133.
 
Event or Problem Description
It was reported that a revision surgery was performed to remove two mobi-c devices because they were locked in a hyperlordotic position in flexion and extension at c3/4 and c4/5.The patient also had dysphagia.The poly core was protruding anteriorly and subsided at c4/5.A corpectomy with fusion was performed at c4 during the revision.This is report one of two for this event.
 
Additional Manufacturer Narrative
If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event or Problem Description
It was reported a revision surgery was performed to remove two mobi-c devices because they were locked in a hyperlordotic position in flexion and extension at c3/4 and c4/5.The patient also had kyphosis that led to dysphagia, but no pain.The poly core was protruding anteriorly and subsided at c4/5.A corpectomy with fusion was performed at c4 during the revision.This is report one of two for this event.
 
Additional Manufacturer Narrative
If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Additional Manufacturer Narrative
Additional information in h6: component, investigation type, findings, and conclusions.Inspection: x-rays were provided which show that both implants are in a hyperlordotic or "fish mouth" position.Migration of the poly core could not be confirmed from the provided x-rays.The devices were sent to a lab for review.The lab provided photos of the devices which show that one of the poly cores is fractured.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the devices were likely conforming when they left zimvie¿s control.Potential root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to unknown patient or surgical factors.Device usage: this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
 
Event or Problem Description
It was reported a revision surgery was performed to remove two mobi-c devices because they were locked in a hyperlordotic position in flexion and extension at c3/4 and c4/5.The patient also had kyphosis that led to dysphagia, but no pain.The poly core was protruding anteriorly and subsided at c4/5.A corpectomy with fusion was performed at c4 during the revision.This is report one of two for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOBI-C IMPLANT "M" STANDARD 13X15 H5 US
Common Device Name
MOBI-C CERVICAL DISC PROSTHESIS
Manufacturer (Section D)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR  10300
Manufacturer (Section G)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR   10300
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key16024005
Report Number3004788213-2022-00132
Device Sequence Number20556531
Product Code MJO
UDI-Device Identifier03662663000000
UDI-Public(01)03662663000000(17)221001(10)5300162
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P110009
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 07/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/01/2022
Device Catalogue NumberMB3355
Device Lot Number5300162
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2022
Supplement Date Received by Manufacturer01/02/2023
02/14/2023
02/14/2023
Initial Report FDA Received Date12/21/2022
Supplement Report FDA Received Date01/27/2023
03/15/2023
07/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/13/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-