AVANOS MEDICAL INC. MICROCUFF SUBGLOTTIC SUCTIONING ENDOTRACHEAL TUBE, 7.0 MM; VAP MICROCUFF ENDOTRACHEAL TUBES
|
Back to Search Results |
|
| Model Number |
13220 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record is in-progress.The sample is reportedly available for this complaint but was not returned all information reasonably known as of (b)(6) 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
It was reported the "pilot balloon's line snapped." there was no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction d4: lot number.The device history record reviewed noted the lot number provided was not correct for the provided stock code in this complaint.The sample was received without the original packaging.Only the balloon pilot line was returned.The rest of the endotracheal tube (et tube) was not received.The size showing on the device was 7.0.During evaluation, the breaking/ separation points were examined under magnification.The break/ separation appeared as if there was a cutting effect since it was slanting.There was a slight tab/ notch at one edge of the separation points, other than that, the separation points were smooth.The incident reported has been confirmed per the sample evaluation.However, as the lot number was not provided further investigation could not be performed.Root cause could not be determined.All information reasonably known as of 07-mar-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The product involved in the report has been returned and is being processed for evaluation.All information reasonably known as of 25-jan-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|