| Model Number |
0279401100 |
| Medical Device Problem Code |
Continuous Firing (1123)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was continuously activating.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was continuously activating.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the reported failure mode was confirmed.Alleged failure: remains continuously activated while the surgeon does not press the button.The probable root cause/s could be a probe short due to a fluid ingress, or an internal leak due to a broken/damaged bond of the suction tubing.The failure mode will be monitored for future reoccurrence.
|
| |
|