Model Number 0279401100 |
Device Problem
Continuous Firing (1123)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/23/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.
|
|
Event Description
|
It was reported that the device was continuously activating.
|
|
Event Description
|
It was reported that the device was continuously activating.
|
|
Manufacturer Narrative
|
The product was returned for investigation and the reported failure mode was confirmed.Alleged failure: remains continuously activated while the surgeon does not press the button.The probable root cause/s could be a probe short due to a fluid ingress, or an internal leak due to a broken/damaged bond of the suction tubing.The failure mode will be monitored for future reoccurrence.
|
|