| Model Number |
2504 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an acl, a tendon anchor did not clip and got broken when implanted to the patch.The broken anchor was removed with a grasper.Surgery resumed after a non-significant delay, with a back-up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H2: additional information: b5.
|
| |
|
Event or Problem Description
|
|
It was reported that during an acl, for a partial rotator cuff repair, a tendon anchor did not clip and got broken when implanted to the patch.The broken anchor was removed with a grasper.Surgery resumed after a non-significant delay, with a back-up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the packing process found that a material certificate of analysis from each raw material supplier is required.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|