Model Number 0279401200 |
Device Problem
Continuous Firing (1123)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/29/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.
|
|
Event Description
|
It was reported that the device continuously activated.
|
|
Manufacturer Narrative
|
Alleged failure: the electrode works discontinuously, and does not stop when pressing the control the failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be the broken/damaged bond of the polyimide and suction tubing observed during the investigation causing fluid ingress to the pcb which can contribute to a short circuit.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
|
Event Description
|
It was reported that the device continuously activated.
|
|