BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24700
Medical Device Problem Code Failure to Deflate (4060)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon deflation issue occurred.The target lesion was located in a shunt in the moderately tortuous vessel.A 5.0mmx100mmx80cm (4f) sterling balloon catheter was advanced for dilation.However, during the procedure, deflation trouble was encountered.The indeflator was used to deflate the balloon in an inflated state.The patient did not have any symptoms while the balloon remained inflated.The catheter was removed with no patient complications nor injuries were reported.
 
Event or Problem Description
It was reported that balloon deflation issue occurred.The target lesion was located in a shunt in the moderately tortuous vessel.A 5.0mmx100mmx80cm (4f) sterling balloon catheter was advanced for dilation.However, during the procedure, deflation trouble was encountered.The indeflator was used to deflate the balloon in an inflated state.The patient did not have any symptoms while the balloon remained inflated.The catheter was removed with no patient complications nor injuries were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a sterling balloon catheter stuck inside a merit 6f sheath.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed that the shaft us stretched 11.2cm and 16.5cm from the tip.Microscopic examination revealed no additional damages.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis found damage to the device that could have contributed to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key16025910
Report Number2124215-2022-53017
Device Sequence Number5643020
Product Code DQY
UDI-Device Identifier08714729122821
UDI-Public08714729122821
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K053116
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/19/2025
Device Model Number24700
Device Catalogue Number24700
Device Lot Number0030134913
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/09/2022
Supplement Date Received by Manufacturer01/11/2023
Initial Report FDA Received Date12/21/2022
Supplement Report FDA Received Date02/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-