U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM A. COOK AUSTRALIA, PTY LTD ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT PROXIMAL BODY; MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM A. COOK AUSTRALIA, PTY LTD ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT PROXIMAL BODY; MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number G32537
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code Obstruction/Occlusion (2422)
Date of Event 12/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A fenestrated [evar] endovascular repair was performed on an 82-year-old female patient with a zenith fenestrated aaa endovascular graft proximal body.The right renal fenestration did not open, even after ballooning (used a cook medical coda balloon) while attempting to cannulate.May be the two middle gold markers had stuck together.After some manipulation the stent eventually deployed, however, lost function to right renal as the graft covered the renal branch.The physician also stated could not get an.014 wire through the right renal branch of the graft.Additional devices implanted during procedure: endologic main body afx gore vbx 6x29 stent in the sma atrium medical icast 6x22 stent in the left renal artery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT PROXIMAL BODY
Common Device Name
MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
WILLIAM A. COOK AUSTRALIA, PTY LTD
95 brandl street
eight mile plains
brisbane
AS 
MDR Report Key16025917
Report Number3005580113-2022-00140
Device Sequence Number5643563
Product Code MIH
UDI-Device Identifier10827002325378
UDI-Public(01)10827002325378(17)251018(10)AC1119801
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/21/2022
Report Date (Section F) 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/18/2025
Device Model NumberG32537
Device Catalogue NumberZFEN-P-2-28-109-R
Device Lot NumberAC1119801
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date12/14/2022
Device Age2 MO
Event Location Hospital
Date Report to Manufacturer12/21/2022
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date12/21/2022
Date Device Manufactured10/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age82 YR
Patient SexFemale
Date Report Sent to FDA12/21/2022
-
-