• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE Back to Search Results
Model Number 6900P
Medical Device Problem Codes Gradient Increase (1270); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
Date of Event 11/29/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned to edwards for evaluation as it remains implanted.Attempts to retrieve additional information is in process.The investigation is still in progress.Therefore; a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
Edwards received information a patient with an unknown edwards surgical valve implanted for an unknown implant duration underwent valve-in-valve procedure due to unknown reasons.A 29mm 9600tfx transcatheter valve was successfully implanted inside the surgical valve.
 
Additional Manufacturer Narrative
The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction.A definitive root cause cannot be conclusively determined; however, patient and/or procedural factors likely caused or contributed.
 
Event or Problem Description
It was learned that a patient with a 29mm 6900p mitral valve underwent a valve-in-valve procedure after an implant duration of 4 years, 8 months due to stenosis.The patient presented with hf and sob.The tmvr was performed with a 29mm 9600tfx transcatheter valve.The patient expired on pod #17.
 
Additional Manufacturer Narrative
The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
irvine, CA 92614
9492506615
MDR Report Key16026069
Report Number2015691-2022-10226
Device Sequence Number7184667
Product Code DYE
UDI-Device Identifier00690103152654
UDI-Public(01)00690103152654(17)210215
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P860057/S016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/15/2021
Device Model Number6900P
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2022
Supplement Date Received by Manufacturer01/24/2023
02/15/2023
Initial Report FDA Received Date12/21/2022
Supplement Report FDA Received Date02/01/2023
02/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/16/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Life Threatening;
Patient Age49 YR
Patient SexFemale
-
-