| Model Number |
VCP269H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint#: (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength ¿ = 275 g/m.(b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: did the needle fall into the patient? if yes, was the needle piece(s) retrieved during the same procedure? what measures were taken to retrieve the broken piece? was there any additional tissue damage as a result of searching for the needle piece? does a piece of the needle remain in the patient¿s tissue? if yes, are any plans in place to remove the needle piece in future? what tissue structure the broken needle was located? what is the surgeon's opinion of consequences to the patient? the tip of the needle was not found and uncertain if it broke off into the patient.The xray was performed per policy to be sure and findings were negative.The magnet was also used and nothing was found.It is possible the needle tip could have fallen on the table or floor and not found/discarded until cleaning of the or room by our environmental services team.This medwatch report is in response to receipt of a report received through fda's medsun program: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a hip surgery on (b)(6) 2022 and suture was used.The tip of the needle broke off while closing the subcutaneous layer of the hip incision.Exploration of the wound performed.Magnet is used on the wound and x-ray was performed.No patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 2/7/2023 this is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device returned.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|