The reason for this revision surgery was reported as an infection.The previous surgery and the surgery detailed in this event occurred 1 month and 3 weeks apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was a nonconformance associated with the main part #952-28-44g, empowr acetabular, dual mobility, poly bearing, id28, od44 which document that out of 20 parts lot 2 parts were rejected and scrapped due to surface finish was not acceptable.All other items in the lot were met with fit, form and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
|