| Model Number |
MLE-24-012 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device referenced in this report has been returned to olympus.The evaluation has not yet been completed.The investigation is ongoing.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.
|
| |
|
Event or Problem Description
|
|
The customer reports during an unspecified procedure using a cutting loop, the cutting loop would not meet connection to cord in working element.The user had to open another.There was no risk to staff or patient.The procedure was prolonged long enough to get new equipment with no further consequences to the patient reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Device evaluation found the visual inspection with no abnormality or thermal damage on the loop, and the shaft is straight.There was one extra brass crimp connector on the shaft proximal end; therefore, the device now has two connectors instead of one as it should be.Minor scratches were found on its surface.In addition, the device was able to produce energy output during further testing.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the reported event was caused by an extra brass crimp got stuck on the proximal end of the electrode which could have caused fitting issue.The following information is stated in the instructions for use (ifu): "no modification of this device/equipment is allowed.Modification of this device/equipment may result in injury to the patient and/or operator, and damage to the equipment and/or peripheral instruments.This device has no serviceable parts." correction: the manufacturer has determined there is no potential for this issue of the cutting loop not connecting to the cord in the working element to cause or contribute to an adverse event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to correct the legal manufacturer¿s contact information and facility registration number.The facility registration number is 3011050570.
|
| |
|