|
As reported for this event by the customer, during a therapeutic laparoscopic lobectomy procedure, the device probe tip broke off and fell into the patient body.The broken piece was immediately removed from the patient body.Thereafter, the procedure was completed normally without any delay with a new similar device from the same lot, which was in stock.There is no harm or adverse impact to the patient.The functional mode at the time of the event was seal and cut mode.The device was inspected prior to use with no abnormality noted.
|
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.The device was returned and an evaluation was completed for it.During inspection, olympus confirmed the reported event, the probe was broken.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the probe broke due to contact with a surgical instrument or the non-insulated area of grasping section.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the event: 1.During output in seal & cut mode, the probe came in contact with hard tissue, metal or a surgical instrument.Consequently, a scratch was made on the probe.2.The probe received an output load in seal & cut mode or received a load when grasping tissue.As a result, the probe cracked from the scratch.3.The probe broke when added load.Or 1.The distal end of the tissue pad was worn away because ¿seal & cut¿ output was activated while grasping nothing (including the case the user kept activating after cutting tissue).2.The tissue pad was worn away, causing the non-insulated of the grasping section to touch the probe.3.¿seal & cut¿ output was activated under this situation, then the scratches indicating that the probe and grasping section were in contact with each other were made.4.The probe received an output load in seal & cut mode or received a load when grasping tissue.As a result, the probe cracked from the scratch.5.The probe broke when added load.The following information is included in the instructions for use (ifu): ¿do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ ¿when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.¿ ¿the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, other instruments, or forceps, and others.Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/split/protruding/partial separating of the tissue pad.In turn, the probe tip may break before displaying an error window or generating an alarm tone.¿ ¿during output, do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments.Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities, which may lead to burns of the tissue by a high-frequency leak current output due to destruction of the insulation structure.¿ olympus will continue to monitor the performance of this device.
|