EXACTECH, INC. EQUINOXE REVERSE COMPONENTS; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Model Number |
320-36-00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Muscle/Tendon Damage (4532); Bursitis (4832)
|
| Date of Event |
11/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported by the equinoxe shoulder study, approximately 9 months post op the initial reverse tsa, this 83 y/o female patient developed pain over the last two weeks.Whenever she got in and out of van, she had to pull herself up to the seats using the right hand multiple times.The case report form indicates this event is possibly related to devices and/or procedure.This event report was received through clinical data collection activities.Action taken: apply ice and heat packs.The outcome is continuing.No additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of the shoulder pain cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition as associated with the interaction between the implanted device and the patient due to patient illness, unique anatomy, or other condition [pulling with her arm on the replacement side] that impacts the performance of the device.It is well known that excessive activity and trauma affect joint replacements.Concomitant medical products: 300-01-10 - equinoxe humeral stem primary press fit 10mm, 320-10-00 - equinoxe reverse tray adapter plate tray +0, 320-31-36 - glenosphere, 36mm, 320-35-01 - small glenoid plate.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: d1, d4(remove all from initial report-suspect device unknown), g4 (remove 501k from initial report-unknown), h6: clinical codes, impact code, type of investigation and investigation findings.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|