| Model Number |
168-390 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/05/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation and exchange of textured device due to patient's concern with product.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ deflation: observed opening on posterior side assessed as fold flaw opening.¿ anxiety-product/procedure: unable to observe as it is a medical event not related to the device.Additional observations: ¿ observed creases on the device.¿ observed wear abrasion on the device.¿ black particles observed in the surface of the shell.¿ observed non penetrating nick on the device.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation and exchange of textured device due to patient's concern with product.Device has been explanted and replaced.
|
| |
|