Investigation: visual inspection: - insertion point on the pediatric prechamber permeability test: the test showed that the progav 2.0 with the pediatric prechamber is permeable.Computer - controlled test: to check the flow rate of the valve, the valve was tested on a miethke computer- controlled testing apparatus and passed the standard test.The flow of the test liquid was reduced step by step from 60 ml/h to 5 ml/h (in accordance with iso 7197).Distilled water was used as the test-liquid.The resulting pressure was measured.The computer- controlled test showed the opening pressure of the progav 2.0, at a reference flow rate of 20 ml/h in a horizontal position of the valve, to be 11,51 cmh2o.This is within the specified tolerance of 13 cmh2o ± 3 cmh2o.Adjustability test: the progav 2.0 was found to be adjustable within the specified range.Braking force and brake function test: the braking force of the progav 2.0 was within the specified tolerance and the brake function operated as expected.Internal inspection of product: after dismantling of the valve, no visible deposits were found in progav 2.0.For more detailed visibility, possible proteins/deposits in progav 2.0 were coloured by using a colouring solution.[colouring solution consisting of coomassi brilliant blue r mixed with 0.6% ethanol and distilled water].After staining, deposits were detected.Results: for the sake of thoroughness and transparency, we would like to point out that an adjustment test was already carried out after the revision at the hospital.Based on our investigation results, we can determine deposits.The deposits had no effect to the technical properties at the time of the investigation.The cause of the aforementioned functional impairment is not known to us at the time of the examination.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.From our point of view, no further regulatory actions are required.
|