No product was returned for evaluation.Radiographs provided confirm the complaint.Patient's bone quality and post-operative physical activity is unknown.It is unknown if the patient suffered a fall.Though a definite root cause cannot be determined, load changes associated with the identified completed fusion and continued use as the likely possible cause though post-operative physical activity may have been a contributor.No additional investigation necessary.Labeling review: "potential adverse events and complications: as with any major surgical procedures, there are risks involved in orthopedic surgery.Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant components"."warnings, cautions and precautions: these devices can break when subjected to the increased load associated with delayed union or nonunion.Correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant.While proper selection can minimize risks, the size and shape of human bones present limitations on the size and strength of implants.Metallic internal fixation devices cannot withstand the activity levels and/or loads equal to those placed on normal, healthy bone.These devices are not designed to withstand the unsupported stress of full weight or load bearing alone.These devices can break when subjected to the increased load associated with delayed union or nonunion.Internal fixation appliances are load-sharing devices that hold bony structures in alignment until healing occurs.If healing is delayed, or does not occur, the implant may eventually loosen, bend, or break.Loads on the device produced by load bearing and by the patient¿s activity level will dictate the longevity of the implant"."patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed"."pre-operative warnings: care should be used during surgical procedures to prevent damage to the devices and injury to the patient"."post-operative warnings: damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration, as well as to other complications"."lifetime of the device: the expected life cycle of implants is determined through astm f1717 testing astm f1717 standard test methods for spinal implant constructs in a vertebrectomy model.5 million cycle endurance limit in astm f1717 represents 2.5 years of in vivo usage per the rationale presented here: the purpose of spinal fusion implants is to provide short term stability until arthrodesis takes place.Following fusion, the device is expected to be unloaded due to load-sharing with the newly formed bone mass.Spinal fusion should occur within 12 months of surgery.According to hedman¸ et al.1991, the average person makes approximately 2 million strides per year 1 million gait cycles.Testing to 5 million cycles is equivalent to a lifetime of approximately 2.5 years of activity, providing a safety factor of 2.5".
|