Event or Problem Description
It was reported that shaft break occurred.A 7.0 x40, 75cm charger balloon catheter was advanced for dilatation; however, during procedure the shaft of the balloon broke and had to be snared out of the body.The procedure was completed and there were no patient complications reported.
 
Event or Problem Description
It was reported that shaft break occurred.A 7.0 x40, 75cm charger balloon catheter was advanced for dilatation; however, during procedure the shaft of the balloon broke and had to be snared out of the body.The procedure was completed and there were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: a charger device 18 atm was received for analysis.A visual examination of the shaft identified an inner balloon shaft break 4mm proximal from the proximal markerband.A visual examination identified complete circumferential tear 5mm distal from the proximal balloon sleeve.A visual examination observed no damage to the tip of the device.A visual examination found no issues or damage to the markerbands of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHARGER
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key16030192
Report Number2124215-2022-51872
Device Sequence Number7475128
Product Code FGE
UDI-Device Identifier08714729804536
UDI-Public08714729804536
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K112697
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/22/2025
Device Model Number24681
Device Catalogue Number24681
Device Lot Number0029642279
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/01/2022
Supplement Date Received by Manufacturer12/21/2022
Initial Report FDA Received Date12/21/2022
Supplement Report FDA Received Date01/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/23/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-