| Model Number |
24681 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 7.0 x40, 75cm charger balloon catheter was advanced for dilatation; however, during procedure the shaft of the balloon broke and had to be snared out of the body.The procedure was completed and there were no patient complications reported.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 7.0 x40, 75cm charger balloon catheter was advanced for dilatation; however, during procedure the shaft of the balloon broke and had to be snared out of the body.The procedure was completed and there were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: a charger device 18 atm was received for analysis.A visual examination of the shaft identified an inner balloon shaft break 4mm proximal from the proximal markerband.A visual examination identified complete circumferential tear 5mm distal from the proximal balloon sleeve.A visual examination observed no damage to the tip of the device.A visual examination found no issues or damage to the markerbands of the device.
|
| |
|