BECTON, DICKINSON & CO., (BD) BD VACUTAINER® SERUM BLOOD COLLECTION TUBE; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 365905
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2022
Event Type  malfunction  
Event Description
It was reported when using the bd vacutainer® serum blood collection tube there was foreign matter in the tube(s) biological and non-biological.The following information was provided by the initial reporter.The customer stated: "a black foreign matter is adhering to the lower part of the cap.".
 
Manufacturer Narrative
Investigation summary: bd received 1 samples and 3 photos for investigation.The photos were reviewed and the customer¿s indicated failure mode for embedded foreign matter was observed.Additionally, the customer samples were evaluated by visual examination and the indicated failure mode for embedded foreign matter with the incident lot was observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode embedded foreign matter.Bd determined that the root cause of the indicated failure mode was attributed to inadequate purging of the line during the injection molding start-up process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SERUM BLOOD COLLECTION TUBE
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16030211
MDR Text Key307998258
Report Number1024879-2022-00734
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number365905
Device Lot Number2161923
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-