| Model Number |
0502539010 |
| Medical Device Problem Code |
Optical Distortion (3000)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foggy image.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: cib, sophia, surgery, water in lens, sjc0015847 [update - moisture could not be wiped away.3 minutes of delay to grab new scope and open it.] the failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be loose objective lens, rough handling during sterilization and/or product transport.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foggy image.
|
| |
|