• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES LLC EDWARDS SAPIEN 3 ULTRA TRANSCATHER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES LLC EDWARDS SAPIEN 3 ULTRA TRANSCATHER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9750TFX
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/01/2022
Event Type  Death  
Event Description
Thv/tvt registry summary reporting for adverse event submission for quarter 4 2022 data extract for aortic death events for the sapien 3 ultra valve.This report summarizes 2 device embolization death events for the sapien 3 ultra transcatheter heart valve.The age range for these events is 81-84 years.The breakdown for gender is as follows: 2 females.
 
Manufacturer Narrative
This report provides data from thv/tvt registry exemption number e2016006 and summarizes 2 device embolization death events for the sapien 3 ultra transcatheter heart valve in the aortic position.The "time to event" (tte, in days) for this event was 0.5.The device identification (di) numbers for edwards sapien 3 ultra transcatheter heart valve are: (b)(4).Per the instructions for use, valve embolization is a known potential adverse event associated with the tavr procedure.There are multiple patient and procedural factors that alone or in combination can cause or contribute to aortic embolization, including improper positioning prior to deployment, poor image intensifier angle, poor coaxial alignment of the valve and delivery system, severe septal hypertrophy, minimally or bulky/severely calcified aortic leaflets, preserved ejection fraction, loss of pacing capture, rapid deployment, release of stored tension during deployment, and movement of the delivery system by the operator.The thv training manuals instruct the operator on proper positioning and deployment of the valve, including all procedural and anatomical considerations.Physicians are extensively trained by edwards before they are qualified to use the thv.Training includes patient screening, device preparation, approach, deployment, imaging, procedure-specific training manuals and proctored procedures.The correct alignment and positioning of the device at the point of deployment is emphasized as a key factor to the placement and fixation of the device.Operators are also instructed to use fluoroscopy as the primary method of visualization for positioning and deployment.In patients with high-risk anatomical features for aortic embolization (i.E.Minimal leaflet calcification, severe septal hypertrophy), bav may provide indication of potential balloon movement during valve deployment.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Specific clinical and procedural details are not available to determine potential contributing factors to the event, or if the event is related to an edwards device.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 ULTRA TRANSCATHER HEART VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key16030700
MDR Text Key305971507
Report Number2015691-2022-09889
Device Sequence Number1
Product Code NPT
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P140031
Exemption NumberE2016006
Number of Events Reported2
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number9750TFX
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/28/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age83 YR
-
-