• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL (SINGAPORE) BD INSYTE¿ IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL (SINGAPORE) BD INSYTE¿ IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Model Number 381234
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/02/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the bd insyte¿ iv catheter experienced breakage requiring surgical intervention.The following information was provided by the initial reporter: upon advancing the ivc, it "scrunched" up and formed a kink.At this point, the ivc had to be removed, so nurse applied pressure on the iv site and pulled out the ivc only to realize that the front bit of the plastic was stuck inside the dog's forearm.Surgeon subsequently operates on the dog to remove the stuck piece.
 
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd insyte¿ iv catheter experienced breakage requiring surgical intervention.The following information was provided by the initial reporter: upon advancing the ivc, it "scrunched" up and formed a kink.At this point, the ivc had to be removed, so nurse applied pressure on the iv site and pulled out the ivc only to realize that the front bit of the plastic was stuck inside the dog's forearm.Surgeon subsequently operates on the dog to remove the stuck piece.
 
Additional Manufacturer Narrative
H6: investigation summary- no photos or samples were received by our quality team for evaluation therefore the failure modes could not be verified.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.Based on the quality team's investigation, the root cause of this incident cannot be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE¿ IV CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16030782
Report Number8041187-2022-00762
Device Sequence Number14863659
Product Code FOZ
UDI-Device Identifier30382903812340
UDI-Public30382903812340
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K151698
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number381234
Device Catalogue Number381234
Device Lot Number2140315
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2022
Supplement Date Received by Manufacturer01/11/2023
Initial Report FDA Received Date12/21/2022
Supplement Report FDA Received Date01/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-