Model Number ORT200
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Preliminary investigation on-site through a third-party representative indicates the malfunction is attributed to the actuator component within the rotational locking mechanism.The manufacturer of the operating room table is working to further investigate the malfunction and collect the actuator component for analysis.
 
Event or Problem Description
It was reported that during a quality assurance/functional check, the ort200 rotational locking mechanism would not lock the table position into place.No patient was involved at the time.
 
Additional Manufacturer Narrative
Investigation of similar events by the supplier of the actuator component has identified reported malfunctions due to a switch within the actuator component.The manufacturer of the operating room table has initiated internal corrective action to further address this issue.Based on event details and isolation of the functional issue at the customer site to the actuator component, this event is within scope of the internal corrective action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMRIS ORT200 OPERATING ROOM TABLE
Common Device Name
OPERATING ROOM TABLE
Manufacturer (Section D)
IMRIS - DEERFIELD IMAGING, INC.
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer (Section G)
IMRIS - DEERFIELD IMAGING, INC.
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer Contact
paul campbell
5101 shady oak road
minnetonka, MN 55343-4100
7632036344
MDR Report Key16031045
Report Number3010326005-2022-00019
Device Sequence Number5643202
Product Code KXJ
UDI-Device Identifier00857534006349
UDI-Public(01)00857534006349
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberORT200
Device Catalogue Number114148-000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2022
Supplement Date Received by Manufacturer11/22/2022
Initial Report FDA Received Date12/21/2022
Supplement Report FDA Received Date03/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/20/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-