| Model Number |
000702 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that post gastrostomy feeding tube placement procedure, the connection of the cap was allegedly broken.It was further reported that the device was removed.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one ponsky dual port feeding tube adapter was received for evaluation.Gross visual, microscopic visual and functional evaluation were performed on the returned device.The investigation is confirmed for the reported fracture and identified material separation issue as a complete break was noted on the closure strap segment.The surface was noted to be granular.A definitive root cause could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that post gastrostomy feeding tube placement procedure, the connection of the cap was allegedly broken.It was further reported that the device was removed.There was no reported patient injury.
|
| |
|