Manufacturers ref# (b)(4) summary of investigational findings: the filter was placed successfully on (b)(6) 2022.A ct follow-up on the (b)(6) 2022 confirmed that a tilt was present.On (b)(6) 022 the patient had open surgery to remove ovarian cancer and it was observed that the filter had penetrated the ivc.The physician who placed the filter confirmed that an approximately 5mm penetration was present.The surgeon made a protective surgical procedure by covering the adjacent structures with fat to avoid further damage from the filter.One fluoroscopic image and several images from a ct scan were provided for imaging review.Per imaging reviewers findings: one fluoroscopic image at time of ivc filter placement, and several images from a ct scan of the abdomen and pelvis with iv contrast, along with the complaint report, were submitted for review.Single image from time of ivc filter placement demonstrates the gunther tulip deployment sheath via jugular approach just below the level of the inflow of the renal veins.Contrast is opacifying the ivc which measures approximately 27 mm in diameter at the infrarenal level.A gunther tulip ivc filter is deployed in the infrarenal location with 0° of tilt present.There is no evidence of immediate penetration and the primary and secondary legs look well aligned.The next 3 images are from a ct scan of the abdomen and pelvis with iv contrast.Two of the images are in the coronal plane, and are of the same slice, one is a magnified view, and the third image is an axial image.In the axial plane, the primary filter legs are located at the 12:00, 3:00, 6:00 and 9:00 regions.The primary filter leg at the 3:00 region extends outside the margin of the ivc and appears to abut the posterior lateral wall of the aorta at approximately the 7 o'clock position.On the coronal plane this filter leg appears to extend approximately 10 mm outside of the expected boundary of the ivc towards the left.There is now also development of 10° rightward tilt of the filter that was not present on the placement images.There are multiple avidly enhancing tumors partially imaged throughout the abdomen.Imaging reviewer´s impressions: per the complaint report, the ivc filter was placed on (b)(6) 2022 for deep vein thrombosis via a jugular approach.There is no evidence of immediate filter tilt or penetration on the single image submitted for review.In addition, the ivc anatomy was considered normal.The complaint report describes development of a filter tilt confirmed on follow-up ct image dated (b)(6) 2022, 5 days after the placement of the ivc filter.The images submitted for review do not contain the date or time stamp to confirm the submitted images were actually acquired on this date.Assuming they were, these images do demonstrate interval development of a 10 degree rightward tilt of the ivc filter with extension of the primary filter leg at the 3:00 region approximately 10 mm outside of the expected wall of the ivc.The ivc filter foot abuts the posterior wall of the aorta, indicating a grade 3 interaction.There are no signs of stranding or haziness adjacent to the filter leg or the aorta.There is no obvious sign of filter fracture.On (b)(6) 2022, patient underwent open abdominal surgery, during which it was noted that the ivc filter leg was penetrating through the wall of the ivc.There was no comment whether there was any bleeding associated with this penetration, or if the ivc filter was interacting with any adjacent structures.However, given this finding, the surgeon packed this region with fat in order to prevent potential future damage.At this point, the cause of penetration, is indeterminate, but likely multifactorial.Conical filter design, dwell time, significant filter tilt, and ivc diameter measuring less than 24 mm have been described as potential contributing factors to the development of ivc filter penetration.In this case, the only confirmed factor known to potentially contribute to development of filter tilt is a conical filter design.However, if this degree of tilt did occur within 5 days, that is very unusual.There is no discussion in the complaint report regarding any predisposing or compounding factors that may have contributed to worsening tilt, such as major abdominal surgery with manipulation of the ivc or endovascular procedures passing devices through the ivc filter.Therefore, the cause of the filter tilt may be related to intrinsic factors with either the filter design and with the patient's anatomy/biology.Furthermore, perforations are well described complication of ivc filters and typically have no clinical relevance.In this case, there was no discussion if the penetration was resulting in any clinical symptoms or had any clinical relevance besides being visualized during surgery.Importantly, if the ivc filter is no longer clinically indicated, it should be removed to avoid any further issues.No evidence to suggest that the product was not manufactured according to specifications.Based on the provided information and imaging review the filter perforation was confirmed.The exact cause for the penetration cannot be established.However, filter perforation is a known adverse event and if the filter is no longer clinically needed it should be removed to avoid further issues.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma.Cook medical will continue to monitor for similar events.This report is required by the fda under 21cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|