• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON, S.A. BD PHASEAL¿ OPTIMA PROTECTOR (P13-O); INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON, S.A. BD PHASEAL¿ OPTIMA PROTECTOR (P13-O); INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 515060
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Chemical Exposure (2570)
Date of Event 12/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medical device expiration date: unknown.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the bd phaseal¿ optima protector (p13-o) leaked while extracting medication from the vial.The following information was provided by the initial reporter: "p-13 leaked during drug extraction from vial".
 
Event or Problem Description
It was reported that the bd phaseal¿ optima protector (p13-o) leaked while extracting medication from the vial.The following information was provided by the initial reporter: "p-13 leaked during drug extraction from vial".
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 06-feb-2023.H6: investigation summary: one protector assembled onto the vial along with the n40-o injector was provided to our quality team for investigation.After proper decontamination, the product was visually inspected, no defects or evidence of improper assembly was observed.Functional evaluations were performed, liquid was able to move from the syringe, into to the vial and back without issue, no leaks were noted.The system was inverted and again no leakages were observed between the protector and vial or any of the connections.Product undergoes inspections throughout the manufacturing process to ensure the quality and functionality of the device.As the lot involved in this incident is unknown, a device history review cannot be performed, and additional retained samples cannot be evaluated.Based on the available information we are not able to identify a root cause at this time.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHASEAL¿ OPTIMA PROTECTOR (P13-O)
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer (Section G)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16032179
Report Number3003152976-2022-00581
Device Sequence Number20244822
Product Code ONB
UDI-Device Identifier00382905150609
UDI-Public00382905150609
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K181221
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number515060
Device Catalogue Number515060
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2022
Supplement Date Received by Manufacturer02/17/2023
Initial Report FDA Received Date12/22/2022
Supplement Report FDA Received Date03/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-