| Model Number |
10379675 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer has been asked to review maintenance and technique and has been asked to run a comparison study to confirm instrument operation.Investigation underway.The cause of this event is unknown.
|
| |
|
Event or Problem Description
|
|
The customer reported that they received several discrepant leukocyte and nitrite results on four patients compared to retesting of different samples on a non-siemens lab instrument.There is no report of injury due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer indicated that instituting proper maintenance resolved their issue.Customer is operational and has stated that they will not provide additional information on this issue.No further investigation possible.
|
| |
|