| Model Number |
115-222 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient reported left side rupture.The device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
No contact information was provided for the initial reporter, therefore additional event, product, and/or patient details are not attainable.The reason for reoperation is: rupture.
|
| |
|
Event or Problem Description
|
|
The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.
|
| |
|
Event or Problem Description
|
|
Patient reported, left side rupture.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ rupture: observed, broken device on posterior side assessed as unidentified (tear) opening (shell thickness was within specification) and missing piece of shell assessed as inconclusive.Additional observations: ¿ red particles on the surface of the shell.¿ crease fold observed.
|
| |
|
Event or Problem Description
|
|
Patient reported left side rupture.The device has been explanted and replaced.
|
| |
|