MEDTRONIC HEART VALVES DIVISION EVOLUT FX DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
Model Number D-EVOLUTFX-34 |
Device Problem
Patient Device Interaction Problem (4001)
|
Patient Problem
Tachycardia (2095)
|
Event Date 12/07/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Product analysis: the device was discarded; therefore, no product analysis can be performed.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Conduction disturbances are known potential adverse effects associated with any cardiac or thoracic procedure (open or catheter-based) and can be resolved with medical treatment or the implant of a permanent pacemaker (with the risk-benefit ratio in favor of implant of the percutaneous aortic valve).A conduction disturbance does not indicate a device malfunction or potential manufacturing issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, upon the first deployment attempt, at 80% deployment, the patient went into ventricular tachycardia.The implanting team were unable to shock the patient out and chose to recapture the valve into the delivery catheter system (dcs) and withdrawn from the patient.Subsequently, bypass was initiated.The patient was stabilized after cardiopulmonary resuscitation was performed.During the second deployment attempt, at 80% deployment, a constrained valve was noted.At this time, it was noted due to the valve constrained; the implanting team considered an infold of the valve frame may have occurred.The valve was recaptured into the dcs and withdrawn from the patient.A new valve was loaded and deployed successfully.No additional adverse patient effects were reported.
|
|
Search Alerts/Recalls
|
|
|