| Model Number |
TG0704509 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Material Too Rigid or Stiff (1544); Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Conclusion not yet available, evaluation in process.A follow-up will be submitted as soon as the investigation is complete.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|
| |
|
Event or Problem Description
|
|
It was reported the 8fr tourguide catheter was used in an extremely tortuous anatomy.It was hard to insert and use in the iliac vessel.The doctor tried to be gentle and careful, however the catheter kinked at the proximal end close to the handle.Eventually the kinked catheter broke, and had to be cut to remove it.The customer reported the event date and lot number as unknown.The reported 8fr tourguide is not a valid model number.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: date of event is 08-dec-2022.The device was used in treatment.No product was provided for analysis.The device will not be returned as it was discarded.No pictures are available regarding this complaint.The device was replaced, and the procedure was successfully completed using another tourguide.It was reported the patients anatomy was very likely the reason why the tourguide failed.The device history records were reviewed to confirm that the device passed all applicable in-process and final inspections.Inspection procedures require any oscor product to pass all in-process and qa final inspection before shipping to the customer.At this time, it is not possible to assign a definitive root cause for the event as reported.The complaint is non-verifiable as the product was not returned for evaluation.No further follow-up is required.No corrective or preventive action resulted after investigation of this event.The event will be re-evaluated if additional information becomes available.Oscor will continue to monitor this event type and risk.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|
| |
|
Event or Problem Description
|
|
Customer provided additional event information.The purpose of the procedure was to place a new stent to the base of left renal artery.The patient had been treated with a fenestrated aortic endograft and the base of the left renal artery had been stented, but the previous stent was suboptimally and there was endoleak.Access was made through right common femoral artery (cfa), which due to previous endovascular operations was extremely scarred.After puncture even though we used amplatz super stiff wire the access had to be enlarged step-by-step with 4 fr, 5 fr, 6 fr and finally 9 fr dilators before 7 fr tourguide could be inserted into the cfa.The tip of the tourguide was brought to the level of the renal arteries through tortuous iliac arteries.The dilator was removed and the tip of the tourguide was arched.After that the tourguide was very gently rotated less than 90 degrees to locate the base of the previous stent in the left renal artery.During the manipulation of the tourguide there was no resistance felt.Suddenly the tip of the tourguide stopped responding to the manipulation.When checked with fluoroscopy the shaft of the tourguide had torn into two, just above the handle.The procedure was completed successfully.However, there was about a 30-minute procedure delay, to exchange to a new tourguide.The anatomy was very likely the reason why the tourguide failed.As mentioned before, the access point in the cfa was extremely scarred and the iliac arteries were tortuous.Even thought the device was discarded, the customer was able to provide the model number tg0704509, and the lot number dp-16192.
|
| |
|