Model Number PLC271400
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 12/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: as gore was unable to determine which device was involved in the event if any; additional device(s) implanted include: plc271000/17381358, udi: (b)(4).W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
On (b)(6) 2018, this patient underwent endovascular treatment for abdominal aortic aneurysm and a right common iliac artery aneurysm and was implanted with gore® excluder® aaa endoprostheses featuring c3® delivery system and gore® excluder® lliac branch endoprosthesis (ibe).The patient tolerated the procedure.On (b)(6) 2022, the physician reported the patient had a distal type i endoleak due to a lack of apposition of the contralateral leg component caused by dilation of the left common iliac artery.The physician intends to re-intervene on the patient, however one has not yet been scheduled.
 
Additional Manufacturer Narrative
The imaging evaluation performed by a clinical imaging specialist showed the following: two sets of post-implantation cta studies are available for evaluation: post-implantation cta¿s dated (b)(6) 2022 and (b)(6) 2023.The (b)(6) 2022 images show a lack of distal limb apposition in the lci.Cannot confirm contrast outside the implanted devices with available images.The images on (b)(6) 2023 show an ibe device and internal iliac component have been implanted in the left common iliac artery, the left internal iliac artery and left external iliac artery.No endoleaks are visualized.
 
Additional Manufacturer Narrative
Corrected date of reintervention.
 
Event or Problem Description
On (b)(6) 2023, the patient underwent endovascular reintervention and the lcia was extended with a gore® excluder® lliac branch endoprosthesis (ceb) and a gore® viabahn® vbx balloon expandable endoprosthesis within the left internal iliac artery.The patient tolerated the procedure and the distal type i endoleak was resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL SILICON VALLEY B/P
2890 de la cruz blvd.
santa clara CA 95050
Manufacturer Contact
laura crawford
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key16035248
Report Number3013164176-2022-01577
Device Sequence Number5629876
Product Code MIH
UDI-Device Identifier00733132618606
UDI-Public00733132618606
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/02/2020
Device Model NumberPLC271400
Device Catalogue NumberPLC271400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2022
Supplement Date Received by Manufacturer12/14/2022
12/14/2022
Initial Report FDA Received Date12/22/2022
Supplement Report FDA Received Date03/29/2023
04/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/03/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention; Hospitalization;
Patient Age77 YR
Patient SexMale
-
-