• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION EVOLUT PRO PLUS VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION EVOLUT PRO PLUS VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number EVPROPLUS-29US
Device Problem Patient Device Interaction Problem (4001)
Patient Problems Cardiac Arrest (1762); Stroke/CVA (1770); High Blood Pressure/ Hypertension (1908); Cardiac Tamponade (2226); Vascular Dissection (3160)
Event Date 12/08/2022
Event Type  Injury  
Event Description
Medtronic received information that during implant of this transcatheter bioprosthetic valve, the patient developed cardiac tamponade and went into cardiac arrest due to an aortic arch dissection.Percutaneous cardiopulmonary support (pcps) was emergently used.The procedure was transitioned to a thoracotomy operation.An ascending aorta replacement was performed and the transcatheter valve was explanted.The operation was completed and the patient left the operating room "without any problems".No additional adverse patient effects were reported.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received that the cause of the aortic arch dissection was noted to be inadequate control of the patients blood pressure which resulted in high blood pressure, which resulted in the dissection.The ascending aorta replacement and valve replacement were noted to be successful and the patients cardiac function was restored.It was reported that the valve and the delivery catheter system (dcs) did not cause or contribute to the dissection.
 
Manufacturer Narrative
Updated data: b2 - outcome attributed to adverse event to include disability b5 - event description h6 - patient code additional codes - to include imf code f1202 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received that following the valve implant, a cerebral infarction was observed and the patient suffered a severe nerve disorder which remained.No treatment was reported.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT PRO PLUS VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16035525
MDR Text Key306010771
Report Number2025587-2022-03688
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P130021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberEVPROPLUS-29US
Device Catalogue NumberEVPROPLUS-29US
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/09/2023
Date Device Manufactured06/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention; Disability;
Patient Age82 YR
Patient SexMale
-
-