MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722LNAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-722LNAP |
Device Problem
Excess Flow or Over-Infusion (1311)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/07/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
Information received by medtronic indicated that the customer reported the pump was not working properly.It was reported that the pump was not working correctly and it didn't figure out the corrections properly and gave her more than it should have.The customer was brittle because of the blood glucose value fluctuations.Troubleshooting was performed and found that the pump did not make correct calculations based on information entered by the customer.No harm requiring medical intervention was reported.The customer will discontinue using the pump and the insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
Additional information has been added which was not included with the initial report.The information has been provided in section b5 with this report.On the event date 07-dec-2022.Customer returned pump for an alleged high bgs.Pump passed stop current, run current, self test, off no power, a21 error test, rewind, basic occlusion, occlusion, prime/a33, excessive no delivery test and displacement test.Pump history download using thds was successful.Carelink downloaded and function properly.Unit was cut/open and inspected no moisture damage or component damage found inside the pump.Test reservoir lock properly inside the reservoir tube.The following were noted during visual inspection: cracked reservoir tube lip and cracked battery tube threads.Unit passed required testing.Unit functioning properly during testing.Pump passed dat.08780.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
The device was returned for the analysis.
|
|