This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and the device evaluation.Please see updates to h4, h6 and h10.The device was returned to olympus for inspection, and the customer's complaint was confirmed.The broken wire was scorched and melted.Additionally, the repair center found the distal end of the coated portion was torn and the distal side of the coated portion was missing approximately 2.0-4.0 mm.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, probable causes for the cutting wire breakage include: 1.The cutting wire where the wire coating was torn came into contact with the distal end of the endoscope when the forceps elevator was raised.2.Under circumstances described above 1, an electric conduction was activated.This caused the cutting wire to become hot instantly at the contact point.As a result, the cutting wire broke.¿ additionally, probable causes for the tears in the distal end coating portion include: 1.It is possible that the slider was slightly pushed causing the cutting wire to deflect.The coated portion of the cutting wire has possibly been rubbed due to the effect of contacting a metal area of the endoscope.2.It can be inferred that some kind of force might have applied to the coated portion of the cutting wire when the device was withdrawn from endoscope after the coated portion of the cutting wire has torn.This might have caused the coated portion of the cutting wire to detach from the cutting wire.As a result, the coated portion was partially missing.However, a final root cause of these events was unable to be identified.The following is included in the instructions for use: ¿since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.When inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.Do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.Be sure that the rear end of the cutting wire is extended from the distal end of the endoscope.In case the cutting wire contacts the forceps elevator, insufficient output or unintended tissue injury may occur.When activating output, set the output mode of the electrosurgical unit to ¿cut¿ or ¿blend¿.Activating output in the ¿coagulation¿ mode could break the cutting wire.Do not activate output when the distal end of the endoscope is too close to or in contact with body cavity tissue.This could burn the tissue and/or damage the endoscope.¿ olympus will continue to monitor field performance for this device.
|
|
During the device evaluation, it was found that the wire was broken, scorched and melted.Additionally, the distal end coating portion was torn and the coating on the knife wire was damaged by the forceps elevator of the endoscope.This report is being submitted for the malfunctions found during the device evaluation.
|