| Model Number |
70415 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/19/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2022, an active 5 month old female patient was cannulated through the ij with a 15 fr crescent ra dual lumen catheter.It was reported sutures were placed around the gold suture ring and the bifurcation.On (b)(6) 2022, after a nurse observed micro bubbles and a fracture to the cannula, the patient was taken to the operating room where the cannula was replaced to the ij.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation found a kink at the fracture.Furture evaluation suggests a suture was tied near the fracture.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received 17-jan-2023, 3011468686-2022-00020 and 3011468686-2022-00022 are the same patient.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated 3011468686-2022-00020 and 3011468686-2022-00022 are the same patient.The patient, confirmed to be 8 months old, with a history of bronchopulmonary dysplasia was treated with ecmo.On (b)(6) 2022, the patient was initially cannulated.On (b)(6) 2022 and (b)(6) 2022, the patient was re-cannulated through the right ij.The patient was not sedated.On (b)(6) 2022, the patient was decannulated.
|
| |
|