• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUMENIS LTD PULSE; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUMENIS LTD PULSE; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number GA-1008944
Medical Device Problem Codes Defective Device (2588); Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that error 425 (#2 footswitch shorted) occurred.The footswitch needs replacement.The procedure was not completed.There was no patient outcome reported.Boston scientific has been unable to obtain additional information regarding the patient outcome to date, despite good faith efforts.This event is being reported for aborted/cancelled procedure with or unknown sedation.
 
Additional Manufacturer Narrative
The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.The reported device performance allegation cannot be confirmed.The event description indicates that the footswitch needs to be replaced.It is likely that the damaged component could have affected the device performance leading to the event experienced by the customer.Based on the information available, since an expected or random component failure was identified without any design or manufacturing issue, the most probable cause of cause traced to component failure was assigned to this investigation.
 
Event or Problem Description
It was reported that error 425 (#2 footswitch shorted) occurred.The footswitch needs replacement.The procedure was not completed.There was no patient outcome reported.Boston scientific has been unable to obtain additional information regarding the patient outcome to date, despite good faith efforts.This event is being reported for aborted/cancelled procedure with or unknown sedation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
LUMENIS LTD
hakidma 6,
yokneam 20692 04
IS  2069204
Manufacturer (Section G)
LUMENIS LTD
hakidma st 6
yokneam 20692 04
IS   2069204
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key16036281
Report Number2124215-2022-53624
Device Sequence Number20196315
Product Code GEX
UDI-Device Identifier07290109145969
UDI-Public07290109145969
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGA-1008944
Device Catalogue NumberGA-1008944
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/13/2022
Supplement Date Received by Manufacturer01/18/2023
Initial Report FDA Received Date12/22/2022
Supplement Report FDA Received Date01/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/12/2016
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-