| Model Number |
E200-33M |
| Medical Device Problem Codes |
Patient-Device Incompatibility (2682); Intravalvular regurgitation (4068)
|
| Health Effect - Clinical Codes |
Endocarditis (1834); Mitral Valve Insufficiency/ Regurgitation (4451)
|
| Date of Event |
11/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on an unknown date, a 33mm epic plus valve was successfully implanted in a patient.On (b)(6) 2022, it was noted that the patient was suffering from device related endocarditis.It was also noted that the patient had mitral regurgitation.The decision was made to explant and replace the 33mm epic plus valve is an unknown size non-abbott device.The patient status was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, a 33mm epic valve was successfully implanted in a patient.On (b)(6) 2022, it was noted that the patient was suffering from device related endocarditis.It was also noted that the patient had mitral regurgitation.The decision was made to explant and replace the 33mm epic valve with an unknown size non-abbott device.The patient status was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of regurgitation secondary to endocarditis was reported.The valve was received for investigation which found that cusps 1 and 2 were torn and contained calcifications.No inflammation was noted.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the tears could not be conclusively determined; however, one of the tears was associated with calcifications.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|
| |
|